DIGITRONIC MICROMETER INSTRUCTIONS
Power
If the micrometer is not used for five minutes the power will automatically shut off
Powering off the micrometer by pressing the "On/Off— Zero key to save the battery when
not in use is recommended
Use a CR2032 battery and replace the battery when the display begins to blur
Remove the battery cap by turning it counterclockwise with a coin or the supplied wrench.
Insert a new battery with (+) side up. Replace the battery cap by turning it clockwise
Data Output
The output interface is a RS-232C
The micrometer can be attached to a PC’s serial port by an SPC cable
To attach the cable, remove the data output cap and insert the cable
When not using the interface, always keep the data output cap in place
Serial Port Information
Baud rate: 1200 Kb/second Parity: none
Stop bits: 2 Data bits: 7
General Specifications
Measuring force: 5-10N
Operating temperature: 0-40°C
Storage temperature: -20-60°C
Precautions
Do not subject the instrument to blows or shock. Do not drop it or apply excessive force
Do not disassemble the instrument
Do not press the keys with a pointed object
Do not use or store the instrument under direct sunlight
Avoid exposing the instrument to strong magnetic fields and high voltage
Use a soft cloth to clean the instrument. Never use organic solvents such as acetone or
benzene to clean
Clean measuring faces before use
If the instrument is to be stored or left unused for extended periods, remove the battery
Troubleshooting
Problem(s) Cause(s) Solution(s)
Measuring data incorrect 1. Dirty measuring faces 1. Clean measuring faces
2. Zero is incorrect 2. Inspect and reset zero
No display on LCD 1. Battery position is incorrect 1. Reset battery
2. Battery is dead 2. Replace with new battery
1. Flickering display 1. Weak battery 1. Replace battery
2. Display is sporadic 2. Weak battery 2. Replace battery
3. Display remains dead 3. Battery position is incorrect 3. Reset battery
1. Display is blurry 1. Weak battery 1. Replace with new battery
2. Output data is incorrect 2. Weak battery 2. Replace with new battery