750 50 0002 03
14
3. After the measuring insert has been removed, the
flatness value (Delta = maximum - minimum) will be
stored and displayed in the buffer (DLT). A new
measurement can than be performed
As log as the measuring insert is in contact with the surface, the Min / Max or Delta mode can be
selected by pressing the
key. The corresponding values will be displayed.
6.6 Zero setting of the display
A zero setting of the display (for surface measurements or on
centerline distances) is made by pressing the Zero key.
6.7 Checking of squareness deviation
The checking of the squareness deviation includes 4 values
as shown on the displayed schema:
z Distance
Squareness deviation
Inclination
— Rectilinearity
Squareness deviation checking is done as follows:
1. Lock the floating probe movement (chromium-plated
screw).
2. Insert the TRIMOS electronic probe with its support (1)
into the probe holder location bore and lock it using the
knob (2).
Z
X
z